Cereulide Analysis by LC-MS/MS
Cereulide detection and targeted toxin analysis in food for manufacturers, quality assurance, laboratories, and authorities
Cereulide is the emetic toxin produced by certain Bacillus cereus strains and may already be pre-formed in food products. For risk assessment, detection of Bacillus cereus alone is therefore not always sufficient. In many cases, targeted cereulide detection is essential to enable robust conclusions regarding product safety, recall prevention, and quality assurance.
ALS offers cereulide analysis in food using LC-MS/MS across a range of food matrices. Testing is performed according to an in-house method based on ISO 18465:2017, using a 13C-labeled internal standard. This approach supports companies in suspected cases, complaints, release decisions, and the evaluation of sensitive product categories such as infant formula.
Your Benefits with ALS Cereulide Analysis
Targeted cereulide detection instead of relying solely on Bacillus cereus testing
LC-MS/MS analysis with high selectivity and sensitivity
In-house method based on ISO 18465:2017
High precision through 13C-labeled internal standard
Fast processing and short turnaround times
Validated food matrices with defined limits of quantification
Our Services
Cereulide Analysis by LC-MS/MS
Cereulide Analysis by LC-MS/MS
ALS determines cereulide in food using LC-MS/MS. The analysis is performed according to an in-house method based on ISO 18465:2017. A 13C-labeled internal standard is used for accurate quantification.
Cereulide is extracted from samples using acetonitrile and subsequently analyzed by LC-MS/MS. This approach enables selective and sensitive determination of the toxin in validated food matrices.
Requirements for Sampling, Shipping, and Storage
Aseptic sampling, sufficient sample quantity, chilled transport, and frozen storage for reliable analytical results
To ensure that cereulide analysis is performed under appropriate conditions, the following requirements should be met:
- Sample amount: at least 50–100 g
- Sampling: aseptic, preferably in sterile bags or containers
- Shipping: ideally chilled
- Storage: frozen until analysis
Validated Food Matrices and Limits of Quantification
Validated cereulide analysis for infant formula, cereal-based foods, and high-fat matrices with defined LOQs
The method has been validated for various food matrices, enabling cereulide detection across different product categories with clearly defined limits of quantification.
Validated Food Matrices and Limits of Quantification (LOQ)
| Matrix | LOQ |
|---|---|
| Infant formula | 0.05 µg/kg |
| Liquid infant formula | 0.025 µg/kg |
| Cereal-based foods (e.g. potatoes, rice, pasta) | 0.05 µg/kg |
| Oil- and fat-rich matrices | 0.1 µg/kg |
Targeted Toxin Analysis for Cereulide
Why direct detection of the emetic Bacillus cereus toxin is critical for food risk assessment
For assessing potential risks in food, it is essential to distinguish between detection of Bacillus cereus and detection of cereulide. While microbiological testing identifies the organism, targeted toxin analysis determines whether the emetic toxin is actually present in the sample.
Particularly in the context of cereulide in food, retain samples, complaints, sensory deviations, or sensitive product categories such as infant formula, microbiological testing alone is often insufficient. Direct cereulide detection provides a robust basis for release decisions, risk assessments, and recall prevention measures.
Laboratory Workflow
Standardized laboratory workflow for fast, precise, and traceable cereulide analysis
The analytical process is designed for reproducibility and timely turnaround:
- Sample receipt and homogenization
Upon arrival at the laboratory, the sample is homogenized. - Addition of internal standard and sample preparation
The sample is spiked with the 13C-labeled internal standard and prepared promptly. - Extraction with acetonitrile
Cereulide is extracted using acetonitrile. - LC-MS/MS measurement and evaluation
The cereulide concentration is determined and evaluated using LC-MS/MS. - Documentation of results
A test report is issued following evaluation.



























